HepoWP P3/P4 Blood Infusion Warming and Pressure System
Features
|
Touch Screen |
8-inch capacitive touch screen; UI design, more user friendly. |
|
All Functions in One |
Integrates five functions: heating, pressurization, speed monitoring, bubble monitoring and liquid stopping. |
|
Stable Pressure and Constant Rate |
Pressure reaches 300mmHg within 25s, pressure deviation ≤ ±3mmHg; Continuous constant rate in the whole process; Blood or medical fluid components not destroyed in pressurization process. |
|
Effective Treatment |
18 types silicone heating profiles; 32.0-42.0℃ adjustable; deviation ≤ 1℃ . |
|
More Advanced and User Friendly |
Access to Central Monitoring System via WIFI; Single and double channel optional; Heating or pressurizing functions can be used simultaneously or separately; 10000mAh rechargeable lithium battery, battery life over 2 hours; Equipped with 12V vehicle power interface for ambulance use. |
|
Accessories |
![]() |
* The specification is subject to change without prior notice, please refer to real products.
Specifications
|
Item |
Specification |
|
Heating Profile Specification |
Inner diameter 3.5, 5, 7mm with length of 50, 60, 90, 100, 120, 140cm, total 18 sizes optional profiles. |
|
Temperature Range |
30.0℃~42.0℃, 86.0℉~107.6℉ |
|
Temperature Deviation |
≤±1.0℃ |
|
Warming Up Time |
AC power supply ≤4 min; |
|
Overheat Protection |
Over 44℃ fuse protection function |
|
Pressurization Mode |
Infusion pressure bag, air squeezed mode |
|
Infusion Pressure Bag Specification |
100ml, 300ml, 500ml, 1000ml, 3000ml optional |
|
Pressure Range |
mmHg: 1~300mmHg; |
|
Pressure Deviation |
≤±3mmHg |
|
Ultimate Output Pressure |
≤310mmHg |
|
Speed Detection Function |
Available, On or Off optional |
|
Bubble Detection Function |
Available, On or Off optional |
|
Running Time |
Available |
|
Alarm |
High Temp, abnormal temperature, high pressure, air bubble, air bubble sensor abnormal, battery ran out, low temp., air leakage, drip sensor abnormal, low battery, battery disconnected, AC Power disconnected, connection failure, no operation |
|
Item |
Specification |
|
Alarm Volume |
20 levels adjustable. |
|
Screen |
8-inch capacitive touch screen, brightness 1 to 100 levels adjustable |
|
Lock Screen Function |
Available, lock screen time 6 levels adjustable |
|
Event Log |
≥10000, exportable via USB port. |
|
Patient Information Setting |
Patient information including patient name, hospitalization number, bed number, age, gender, height, weight, department, clear patient information, etc |
|
Network |
Access to HepoMS M1 Central Monitoring System via WiFi connection |
|
Battery |
10000Amh rechargeable lithium battery, operating time ≥ 2h (P3 ≥ 3h) |
|
AC Power |
100V-240V~, 50/60Hz. power: 180VA (P3: 120VA) |
|
Direct Dower |
12V (vehicle power supply) |
|
Degree of Protection Against Ingress of Liquids |
IP34 |
|
Dimension |
P4: 230mm×140mm×310mm P3: 200mm×140mm×310mm |
|
Net Weight |
P3: 7.5kg, P4: 8.0kg |
|
Others |
Heating or pressurizing functions can be used simultaneously or separately |